AC inrush current (max.) |
СOLD STAR 30A/115VAC 50A/230VAC |
Over Load Protection |
115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed |
Over voltage protection |
3.8~4.5V 5.75~6.8V 8.6~11.3V 13.8~16.2V 17.2~20.3V 27.6~32.4V 55.2~64.8V
Protection type : Shut down o/p voltage, re-power on to recover |
Leakage current |
Touch current< 100 A/264VAC |
Working temperature |
-30 ~ +70℃(Refer to "Derating Curve") |
Safety standards |
IEC60601-1, TUV EN60601-1, EAC TP TC 004,UL ANSI / AAMI ES60601-1 (3.1 version),
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to EN60335-1 |
ESD |
EN61000-4-2 |
Level 4, 15KV air ; Level 4, 8KV contact |
RF field susceptibility |
EN61000-4-3 |
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz ) |
Surge susceptibility |
EN61000-4-5 |
Level 4, 4KV/Line-FG; 2KV/Line-Line |
Voltage dip, interruption |
EN61000-4-11 |
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods |
Packing |
0.11Kg; 120pcs/14.2Kg/0.97CUFT |