AC DC Power Supply > Medical Power Supply
> Open Frame Type > Single output > 500W > RPS-500 |
|
Features
: |
* 5"×2" miniature size
* 320W convection, 500W force air
* 550W peak power (3sec.)
* Medical safety approved (2 x MOPP) accroding to ANSI/AAMI ES60601-1 and IEC/EN60601-1
* EMI for both Class I & Class II configuration
* -30~+70
℃ wide range operating temperature
* No load power consumption<0.5W by PS_ON control
* High efficiency up to 94%
* Protections: Short circuit/ Overload/ Over voltage/ Over temperature
* 5Vdc standby output, 12Vdc fan
supply, Power Good, Power Fail and remote sense
* Operating altitude up to 4000 meters (Note.5)
* LED indicator for power on
* 3 years warranty
|
Model No. |
RPS-500 |
|
AC input voltage
range |
80 ~ 264VAC 113 ~ 370VDC |
AC inrush current (max.) |
COLD START 40A/115VAC 80A/230VAC |
Over Load Protection |
105 ~ 135% rated output powe
Protection type : Hiccup mode, recovers automatically after fault condition is removed |
Over voltage protection |
13.2 ~ 15.6V, 16.5 ~ 19.5V,19.8V~23.4V, 26.4 ~ 31.2V, 29.7 ~ 35.1V,39.6~46.8V, 52.8 ~ 62.4V
Protection type : Shut down o/p voltage, re-power on to recover |
Leakage current |
Earth leakage current <220 uA/264VAC 50Hz , Touch current < 100 uA/264VAC |
Withstand voltage |
I/P-O/P:4KVAC
I/P-FG:2KVAC
O/P-FG:1.5KVAC |
Working temperature |
-30 ~ +70℃(Refer to "Derating Curve") |
Safety standards |
IEC60601-1, TUV EN60601-1,UL ANSI / AAMI ES60601-1 (3.1 version),EAC TP TC 020.
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved;
Type
M’MENT 1.03CUFT 0.77CUFT 0.86CUFT 0.91CUFT
Design refer to EN60335-1 (By request) |
EMC EMISSION |
Parameter |
Standard |
Test Level / Note |
Conducted emission |
EN55011 (CISPR11) |
ClassⅠ: Class B , ClassⅡ: Class A |
Radiated emission |
EN55011 (CISPR11) |
Class A |
Harmonic current |
EN61000-3-2 |
Class A |
Voltage flicker |
EN61000-3-3 |
---- |
EMC IMMUNITY |
EN60601-1-2 |
Parameter |
Standard |
Test Level / Note |
ESD |
EN61000-4-2 |
Level 4, 15KV air ; Level 4, 8KV contact |
RF field susceptibility |
EN61000-4-3 |
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz ) |
EFT bursts |
EN61000-4-4 |
Level 3, 2KV |
Surge susceptibility |
EN61000-4-5 |
Level 4, 4KV/Line-FG; 2KV/Line-Line |
Conducted susceptibility |
EN61000-4-6 |
Level 3, 10V |
Magnetic field immunity |
EN61000-4-8 |
Level 4, 30A/m |
Voltage dip, interruption |
EN61000-4-11 |
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods |
DIMENSION |
RPS-500:127x76.2x41mm
RPS-500-C:130x86x43mm
RPS-500-TF:130x86x66.5mm
RPS-500-SF:160x86x43mm |
Packing |
RPS-500:0.46Kg,
30pcs,
14.8Kg, 1.03CUFT
RPS-500-C:0.54Kg,
24pcs,
14Kg, 0.77CUFT
RPS-500-TF:0.58Kg,
24pcs,
14.9Kg, 0.86CUFT
RPS-500-SF:0.64Kg,
24pcs,
16.4Kg, 0.91CUFT |
|
Model No. |
DC VOLTAGE |
RATED CURRENT
|
RATED POWER
|
Tol. |
R & N |
Effi. |
Blank |
-C |
-TF/SF |
Blank |
-C |
-TF/SF |
25CFM |
Convection |
25CFM |
Convection |
Built-in fan |
25CFM |
Convection |
25CFM |
Convection |
Built-in fan |
RPS-500-12□ |
12V |
41.6A |
26.7A |
41.6A |
25.8A |
41.6A |
499.2W |
320.4W |
499.2W |
309.6W |
499.2W |
±3% |
200mV |
91% |
RPS-500-15□ |
15V |
33.3A |
21.3A |
33.3A |
20.7A |
33.3A |
499.5W |
319.5W |
499.5W |
310.5W |
499.5W |
±3% |
200mV |
92% |
RPS-500-18□ |
18V |
27.8A |
17.8A |
27.8A |
17.2A |
27.8A |
500.4W |
320.4W |
500.4W |
309.6W |
500.4W |
±3% |
200mV |
92.5% |
RPS-500- 24□ |
24V |
20.8A |
13.4A |
20.8A |
12.9A |
20.8A |
499.2W |
321.6W |
499.2W |
309.6W |
499.2W |
±2% |
200mV |
93% |
RPS-500- 27□ |
27V |
18.5A |
11.9A |
18.5A |
11.5A |
18.5A |
499.5W |
321.3W |
499.5W |
310.5W |
499.5W |
±2% |
200mV |
93.5% |
RPS-500- 36□ |
36V |
13.9A |
8.9A |
13.9A |
8.6A |
13.9A |
500.4W |
320.4W |
500.4W |
309.6W |
500.4W |
±1% |
200mV |
94% |
RPS-500- 48□ |
48V |
10.4A |
6.7A |
10.4A |
6.5A |
10.4A |
499.2W |
321.6W |
499.2W |
312W |
499.2W |
±1% |
200mV |
94% |
|
|