AC inrush current (max.) |
COLD START 35A/115VAC 70A/230VAC |
Over Load Protection |
105% ~ 135% rated
output power
Protection type : Hiccup mode, recovers automatically
after fault condition is removed |
Over voltage protection |
13.5 ~ 15V 16.2 ~ 18.5V 26 ~ 30V 29.5 ~ 33.5V 52 ~ 59.5V
Protection type : Shut down o/p voltage, re-power on to recover |
Over Temperature protection |
Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down
Protection type : (TSW2)Shut down o/p voltage, re-power on to recover
|
Leakage current |
Earth leakage current <250uA / 264VAC, Touch current <100uA/264VAC |
Withstand voltage |
I/P-O/P: 4KVAC
I/P-FG: 2KVAC
O/P-FG: 1.5KVAC |
Working temperature |
-30°C ~ 70°C (refer to output derating curve) |
Safety standards |
IEC60601-1, TUV EN60601-1,EAC TP TC 004,
UL ANSI / AAMI ES60601-1 (3.1 version),
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved;
Design refer to EN60335-1 |
ESD |
EN61000-4-2 |
Level 4, 15KV air ; Level 4, 8KV contact |
RF field susceptibility |
EN61000-4-3 |
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz ) |
Surge susceptibility |
EN61000-4-5 |
Level 4, 4KV/Line-FG; 2KV/Line-Line |
Voltage dip, interruption |
EN61000-4-11 |
100% dip 1 periods, 30% dip 25 periods,100% interruptions 250 periods |
DIMENSION |
PCB type:127*76.2*35mm or 5″*3″*1.37″inch
Enclosed type:130*86*43mm or 5.11″*3.39″*1.69″inch |
Packing |
PCB type:0.37Kg; 36pcs/14.3Kg/1.03CUFT
Enclosed type:0.563Kg; 24pcs/14.5Kg/0.77CUFT |