Power Supply RPS-200
Features :

* 4"×2" miniature size
* Medical safety approved (2 x MOPP) according to ANSI/ AAMI ES60601-1 and IEC/EN60601-1
* Suitable for BF application with appropriate system consideration
* 140W convection, 150W force air
* EMI Conduction for Class B Radiation for Class B with FG (Class I) and Class A without FG (Class II)
* No load power consumption < 0.5W
* Extremely low leakage current
* 12V/0.5A fan supply
* Protections: Short circuit/ Overload/ Over voltage/ Over temperature
* Lifetime > 65K hours
* Operating altitude up to 5000 meters
* 3 years warranty

Model No. RPS-200
AC input voltage range 80 ~ 264VAC 113 ~ 370VDC
AC inrush current (max.) COLD START 30A/115VAC 60A/230VAC
Over Load Protection 110 ~ 140% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
Over voltage protection 13.2 ~ 15.6V, 16.5 ~ 19.5V, 26.4 ~ 31.2V, 29.7 ~ 35V, 52.8 ~ 62.4V
Protection type : Shut down o/p voltage, re-power on to recover
Leakage current Earth leakage current < 130 μA/264VAC , Touch current < 40 μA/264VAC
Withstand voltage I/P-O/P:4KVAC
I/P-FG:2KVAC
O/P-FG:1.5KVAC
Working temperature -30 ~ +70℃(Refer to "Derating Curve")
Safety standards IEC60601-1, TUV EN60601-1, EAC TP TC 004,UL ANSI / AAMI ES60601-1 (3.1 version),
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to EN60335-1
EMC EMISSION Parameter Standard Test Level / Note
Conducted emission EN55011 (CISPR11) Class B
Radiated emission EN55011 (CISPR11) Class A (for Class II);Class B (for Class I)
Harmonic current EN61000-3-2 Class A
Voltage flicker EN61000-3-3 ----
EMC IMMUNITY EN60601-1-2
Parameter Standard Test Level / Note
ESD EN61000-4-2 Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility EN61000-4-3 Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts EN61000-4-4 Level 3, 2KV
Surge susceptibility EN61000-4-5 Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility EN61000-4-6 Level 3, 10V
Magnetic field immunity EN61000-4-8 Level 4, 30A/m
Voltage dip, interruption EN61000-4-11 95% dip 1 periods, 30% dip 25 periods,
95% interruptions 250 periods
DIMENSION PCB:101.6*50.8*29mm or 4″*2″*1.14″inch ; Enclosed type:103.4*62*40mm or 4.07″*2.44″*1.57″inch
Packing PCB:0.19Kg; 72pcs/14.7Kg/0.84CUFT ; Enclosed type:0.3Kg; 60pcs/19Kg/1.06CUFT
Model No. DC VOLTAGE RATED CURRENT
RATED POWER Tol. R & N Effi.
10CFM Convection 10CFM Convection
RPS-200-12□ 12V 16.7A 11.7A 200.4W 140.4W ±2% 100mV 93%
RPS-200-15□ 15V 13.4A 9.4A 201W 141W ±2% 100mV 93.5%
RPS-200- 24□ 24V 8.4A 5.9A 201.6W 141.6W ±1% 120mV 94%
RPS-200- 27□ 27V 7.5A 5.3A 202.5W 143.1W ±1% 120mV 94%
RPS-200- 48□ 48V 4.2A 3A 201.6W 144W ±1% 120mV 95%
 