Power Supply RPS-120
Features : * 3"×2" miniature size
* Medical safety approved (2 x MOPP) according to ANSI/ AAMI ES60601-1 and IEC/EN60601-1
* Suitable for BF application with appropriate system consideration
* 120W convection, 150W force air
* EMI for voth Class I & Class II configuration
* -30~+85℃wide range operating temperature
* No load power consumption < 0.3W
* Extremely low leakage current
* Protections: Short circuit/ Overload/ Over voltage/ Over temperature
* Operating altitude up to 4000 meters
* 3 years warranty
Model No. RPS-120S
AC input voltage range 80 ~ 264VAC 113 ~ 370VDC
AC inrush current (max.) COLD START 30A/115VAC 60A/230VAC
Over Load Protection 130~160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
Over voltage protection 13.2 ~ 15.6V 16.5 ~ 19.5V 26.4 ~ 31.2V 29.7 ~ 35V 52.8 ~ 62.4V
Protection type : Shut down o/p voltage, re-power on to recover
Leakage current Earth leakage current < 150 μA/264VAC , Touch current < 80 μA/264VAC
Withstand voltage I/P-O/P:4KVAC
I/P-FG:2KVAC
O/P-FG:1.5KVAC
Working temperature -30 ~ +85℃(Refer to "Derating Curve")
Safety standards IEC60601-1, TUV EN60601-1, EAC TP TC 004, UL ANSI / AAMI ES60601-1 (3.1 version),
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to EN60335-1 (By request)
EMC EMISSION Parameter Standard Test Level / Note
Conducted emission EN55011 (CISPR11) Class B
Radiated emission EN55011 (CISPR11) ClassⅠ: Class B , ClassⅡ: Class A
Harmonic current EN61000-3-2 Class A
Voltage flicker EN61000-3-3 ----
EMC IMMUNITY EN60601-1-2
Parameter Standard Test Level / Note
ESD EN61000-4-2 Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility EN61000-4-3 Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts EN61000-4-4 Level 3, 2KV
Surge susceptibility EN61000-4-5 Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility EN61000-4-6 Level 3, 10V
Magnetic field immunity EN61000-4-8 Level 4, 30A/m
Voltage dip, interruption EN61000-4-11 95% dip 1 periods, 30% dip 25 periods,
95% interruptions 250 periods
DIMENSION 76.2*50.8*28mm or 3" * 2" *1.1" inch
Packing 0.13Kg; 100pcs/14Kg/1.13CUFT
Model No. DC VOLTAGE CURRENT RATED POWER Tol. R & N Effi.
10CFM Convection 10CFM Convection
RPS-120S-12 12V 11.8A 9.5A

141.6W

114W ±2% 100mV 91%
RPS-120S-15 15V 9.5A 7.6A 142.5W 114W ±2% 120mV 92%
RPS-120S-24 24V 6.25A 5A 150W 120W ±1% 150mV 93%
RPS-120S-27 27V 5.55A 4.44A 149.8W 119.9W ±1% 150mV 94%
RPS-120S-48 48V 3.125A 2.5A 150W 120W ±1% 200mV 93.5%
 