Power Supply RPD-75
Features : ‧4"×2" compact size
‧Medical safety approved (2 x MOPP) accroding to ANSI/AAMI ES60601-1 and IEC/EN60601-1
‧Suitable for BF application with appropriate system consideration
‧Cooling by free air convection
‧EMI class B for class Ⅰ configuration
‧Extremely low leakage current
‧Protections: Short circuit / Overload / Over voltage
‧3 years warranty
Model No. RPD-60
AC input voltage range 90 ~ 264VAC 127 ~ 370VDC
AC inrush current (max.) COLD START 60A/230VAC 30A/115VAC
OVER VOLTAGE CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
Over Load Protection 115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
Leakage current Earth leakage current < 150 A/264VAC , Touch current < 100 A/264VAC
SETUP, RISE TIME 300ms, 15ms/230VAC 300ms, 15ms/115VAC at full load
Withstand voltage I/P-O/P: 4KVAC,
I/P-FG: 1.5KVAC, 1 minute
Working temperature -20 ~ +65°C (Refer to "Derating Curve")
Safety standards UL60950-1,TUV EN60950-1,IEC60601-1, EAC TP TC 004, UL ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1:14 - Edition 3
approved,TUV EN60601-1 approved
EMC standards Parameter Standard Test Level / Note
Conducted emission EN55011 (CISPR11) Class B
Radiated emission EN55011 (CISPR11) Class B
Harmonic current EN61000-3-2 Class A
Voltage flicker EN61000-3-3 -----
EMC IMMUNITY EN60601-1-2
Parameter Standard Test Level / Note
ESD EN61000-4-2 Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility EN61000-4-3 Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts EN61000-4-4 Level 3, 2KV
Surge susceptibility EN61000-4-5 Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted susceptibility EN61000-4-6 Level 3, 10V
Magnetic field immunity EN61000-4-8 Level 4, 30A/m
Voltage dip, interruption EN61000-4-11 100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
MTBF 677.8K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (L*W*H) 101.6*50.8*29mm or 4" * 2" *1.14" inch
PACKING 0.15Kg; 96pcs/15.4Kg/0.89CUFT
RPD-60/ 60W Dual Output Medical Grade
Model No. Output Tol. R & N Effi.
RPD-75A 5V, 0.5 ~ 5.5A
12V, 0.1 ~ 2.2A
±1.5%
±2.0%
80mVp-p 78%
RPD-75B 5V,0.5 ~ 3.85A
24V, 0.1 ~ 1.65A
±1.5%
±2.0%
80mVp-p
100mVp-p
82%
 