AC inrush current (max.) |
10 A(rms)for 115 VAC
5 A(rms)for 230 VAC |
Over voltage protection |
Provided on output #1 only; set
at 115-140%of its nominal
output voltage |
Over temp. protection |
All outputs +/-0.04% /°C maximum |
Withstand voltage |
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground |
EMC standards |
EN55011: Class B conducted, Class A radiated
EN61000-3-2: Harmonic distortion, Class A and D
EN61000-3-3: Line flicker
EN61000-4-2: ESD, +/-8 KV air and +/-6 KV contact
EN61000-4-3: Radiated immunity, 3 V/m
EN61000-4-4: Fast transient/burst, +/-2 KV
EN61000-4-5: Surge,+/-1 KV diff., +/-2 KV com.
EN61000-4-6: Conducted immunity, 3 Vrms
EN61000-4-8: Magnetic field immunity, 3 A/m
EN61000-4-11: Voltage dips, 30% reduction for 500
ms, 60% reduction for 100 ms and>95% reduction
for 10 ms |