AC inrush current (max.) |
3.20 A (rms) for 115 VAC
1.60 A (rms) for 230 VAC |
Over voltage protection |
Provided on output #1 only; set at 112-132%
of its nominal output voltage |
Over temp. protection |
All outputs ±0.04% /°C maximum |
Withstand voltage |
5600 VDC from input to output,
2100 VDC from input to ground,
700 VDC from output to ground
(To verify AC strength, get correct test method
to avoid power supply damage.) |
MTBF |
350,000 hours at full load at 25℃ ambient,
calculated per MIL-HDBK-217F |
EMC standards |
EN55011/EN55022: Class B conducted, class B radiated
EN61000-3-2: Harmonic distortion, class A and D
EN61000-3-3: Line flicker
EN61000-4-2: ESD, ±8 KV air and ±6 KV contact
EN61000-4-3: Radiated immunity, 3 V/m
EN61000-4-4: Fast transient/burst, ±2 KV
EN61000-4-5: Surge, ±1 KV diff., ±2 KV com.
EN61000-4-6: Conducted immunity, 3 Vrms
EN61000-4-8: Magnetic field immunity, 3 A/m
EN61000-4-11: Voltage dip immunity, 30% reduction for
500 ms, 60% reduction for 100 ms, and
>95% reduction for 10 ms |