AC inrush current (max.) |
3.20 A (rms) for 115 VAC
1.80 A (rms) for 230 VAC |
Over voltage protection |
Provided on output #1 only;
set at 112-132% of its nominal output
voltage |
Over temp. protection |
All outputs +/-0.04/°C maximum |
Withstand voltage |
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground |
EMC standards |
EN55011: Class B conducted, class B radiated
FCC: Class B conducted, class B radiated
VCCI: Class B conducted, class B radiated
EN61000-3-2: Harmonic distortion, class A
EN61000-3-3: Line flicker
EN61000-4-2: ESD, +/-8 KV air and +/-6 KV contact
EN61000-4-3: Radiated immunity, 3 V/m @ 80-2500
MHz
EN61000-4-4: Fast transient/burst, +/-2 KV
EN61000-4-5: Surge, +/-1 KV diff., +/-2 KV com.
EN61000-4-6: Conducted immunity, 3 Vrms
EN61000-4-8: Magnetic field immunity, 3 A/m
EN61000-4-11: Voltage dips, 30% reduction for 500
ms,
60% reduction for 100 ms and >95%
reduction for 10 ms |