DC POWER FACTOR (Typ.) |
PF>0.95/230VAC PF>0.99/115VAC at full load |
Leakage current |
Earth leakage current < 300µ A/264VAC
, Touch leakage current < 100µ A/264VAC |
Over Load Protection |
105 ~ 135% rated output power
Protection type : Constant current limiting, recovers
automatically after fault condition is removed |
Over voltage protection |
3.96~67.2V
Protection type : Shut down o/p voltage, re-power
on to recover |
Over Temperature Protection
|
Shut down o/p voltage, recovers automatically
after temperature goes down |
Working temperature |
-40 ~ +70°C (Refer to "Derating Curve") |
Withstand voltage |
I/P-O/P:4KVAC
I/P-FG:2KVAC
O/P-FG:0.5KVAC |
Safety standards |
ANSI/AAMI ES60601-1, IEC60601-1 approved |
EMC Emission |
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3 |
EMC Immunity |
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2 |
Packing |
0.77Kg; 18pcs/14.9Kg/0.9CUFT |