Power Supply MPQ-200
Features : * Universal AC input / Full range
* Low leakage current <180 A
* Protections: Short circuit / Overload / Over voltage / Over temperature
* Free air convection for 140W and forced air convection for 200W
* ANSI/AAMI ES60601-1 medical safety approved
* With power good and fail signal output
* Built-in remote ON-OFF control
* Built-in remote sense function
* Fixed switching frequency at 100KHz
* 3 years warranty
Model No. MPQ-200
AC input voltage range 90~264 VAC, 47~440Hz / 127~370VDC.
AC inrush current (max.) Maximum input current 3.5A at 115VAC, 60Hz or
1.6A at 230VAC, 60Hz with 100% output load.
DC Adjustment Range DC output voltage (or CH1 of multiple output models)
can be adjusted between -5%~+10% rated output
voltage by potential meter.
Over Load Protection Fully protected against short circuit and output overload.
The hiccup type protection will be activated at 120~160%
rated load and recovers automatically after fault condition
is removed.
Over voltage protection Provided on output channel 1 only at 115%~135%
rated output voltage.
Output will be shut down when this protection is activated.
Leakage current Leakage current is less than 180 A at 264VAC for
earth leakage current
Leakage current is less than 100 A at 264VAC for
patient leakage current
Withstand voltage 4000VAC between input and output
1500VAC between input and F.G.
1500VAC between output and F.G.
Working temperature Whole series can operate from -20~70 . Please refer
to the derating curves.
Safety standards Medical: ANSI/AAMIES60601-1, TUV EN60601-1, IEC60601-1 approved
Commercial : Also design refer to UL60950-1, TUV EN60950-1
EMI COMPLIANCE

EMI Specifications: Compliance Level
Conducted & Radiation: EN55011, Class B ; EN55022, Class B
Harmonic distortion: EN61000-3-2
Voltage flicker: EN61000-3-3

EMC standards EMS Specification: Compliance Level
ESD air: EN61000-4-2, Level 3, 8KV
ESD contact: EN61000-4-2, Level 2, 4KV
RF field susceptibility: EN61000-4-3, Level 2, 3V/m
-------------------------------------------Level 3, 10V/m
EFT(Electrical Fast Transient)/Burst: EN61000-4-4,Level 2, 1KV/5KHz
-----------------------------------------------------------Level 3, 2KV/5KHz
Lightning/Surge: EN61000-4-5,Level 4, 2KV/Line-Line
---------------------------------------------4KV/Line-Earth
Conducted RF susceptibility: EN61000-4-6, Level 2, 3Vrms/m
--------------------------------------------------Level 3, 10Vrms/m
Magnetic field immunity: EN61000-4-8, Level 2, 3A/m
----------------------------------------------Level 3, 10A/m
Voltage dip, interruption: EN61000-4-11, Compliance
Digital phone carrier immunity: ENV50204, Level 2, 3V/m, 900MHz
--------------------------------------------------Level 3, 10A/m, 900MHz
DIMENSION 177.8x107.2x35.5mm or 7"x4.22"x1.4"(L*W*H)
Packing 0.66Kg; 24pcs/16.8Kg/0.99CUFT
Model No. Output VOLTAGE RATED CURRENT OUTPUT CURRENT (A) Tol. R & N Effi.
MINIMUM LOAD CONVECTION (max.) WITH FAN (25CFM) PEAK LOAD WITH 25CFM FAN
MPQ-200B 5V 15A 3 12 15 18

±2%

80m

78%
12V 7A 0.7 5.3 7 8.4 ±8% 120mV
-5V 2A 0 1 2 2.4 ±5% 80mV
-12V 2A 0 1 2 2.4 ±5% 80mV
MPQ-200C 5V 15A 3 12 15 18 ±2%

80mV

78%
15V 5A 0.5 4 5 6 ±6% 150mV
-5V 2A 0 1 2 2.4 ±5% 80mV
-15V 2A 0 1 2 2.4 ±5% 80mV
MPQ-200D 5V 15A 3 12 15 18 ±2%

80mV

79%
24V 3A 0.3 2.3 3 3.6 ±8% 180mV
12V 2A 0 1 2 2.4 ±5% 80mV
-12V 2A 0 1 2 2.4 ±5% 80mV
MPQ-200F 5V 15A 3 12 15 18 ±2%

80mV

81%
24V 2.7A 0.3 2.1 2.7 3.3 ±8% 180mV
15V 2A 0 1 2 2.4 ±5% 80mV
-15V 2A 0 1 2 2.4 ±5% 80mV
 
Download PDF Here