E-Star Power Supply Mail to Us power supply products
| AC DC Power Supply | | Medical Power Supply | | DC DC Power Supply | | AC POWER CORD | - | Contact Us | - | Home |
 
AC DC Power Supply
> Medical Power Supply
> Open Frame Type

> 30W

> MFM-30

 
 
 
 
AC DC Power Supply > Medical Power Supply > Open Frame Type > 30W > MFM-30
Power Supply MFM-30
Features : * 2.58"x1.38" compact size
* Medical safety approved (2xMOPP) according to ANSI/ AAMI ES60601-1 and IEC/EN60601-1
* Suitable for BF application with appropriate system consideration
* No load power consumption<0.075W
* Extremely low leakage current
* Wide operating temp range -40~+85°C
* EMI class B for cless II configuration
* Protections: Short circuit/ Overload/ Over voltage/ Over temperature
* No minimum load required
* 3 years warranty
Model No. MFM-30
AC input voltage range 80~264VAC
Frequency Range 47~63Hz
AC inrush current (max.) Cold start 25A/115VAC 45A/230VAC
Over Load Protection 115%~165% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is removed
Leakage current Touch current <80uA/264VAC
Over voltage protection 3.5~4.5V, 5.3~6.8V, 12.6~16.2V, 15.8~20.3V, 25.2~32.4V, 50.4~64V
Protection type: Shut off o/p voltage, re-power on to recover
Working temperature -40°C ~ +85°C (refer to derating curve)
Safety standards IEC60601-1, EN60601-1, EAC TP TC 004, UL ANSI/ AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved; Design refer to EN60335-1
EMC Emission Parameter Standard Test Level / Note
Conducted EN55011 (CISPR11) Class B
Radiated EN55011 (CISPR11) Class B
Harmonic Current EN61000-3-2 Class A
Voltage Flicker EN61000-3-3 -----
EMC Immunity EN60601-1-2
Parameter Standard Test Level / Note
ESD EN61000-4-2 Level 4, 15KV air; Level 4, 8KV contact
RF field susceptibility EN61000-4-3 Level 3, 10V/m(80MHz~2.7GHz)
Table 9, 9~28V/m(385MHz~5.78GHz)
EFT bursts EN61000-4-4 Level 3, 2KV
Surge susceptibility EN61000-4-5 Level 3, 1KV/Line-line
Conducted susceptibility EN61000-4-6 Level 3, 10V
Magnetic field immunity EN61000-4-8 Level 4, 30A/m
Voltage dip, interruption EN61000-4-11 100% dip 1 periods, 30% dip25periods, 100% interruptions 250 periods
Withstand voltage I/P-O/P: 4.0KVAC
MTBF 779Khrs min. MIL-HDBK-217F(25°C)
Dimension 65.5*35.8*23mm(L*W*H)
Packing 0.053kg; 144pcs / 8.6Kg / 0.97CUFT
MFM-30 / 30W High Reliable Green Medical On Board Type
Model No. DC Voltage Rated current Rated power R & N Voltage Tol. Effi.
MFM-30-3.3 3.3V 6A 19.8W 80mVp-p ±2% 82.5%
MFM-30-5 5V 6A 30W 80mVp-p ±2% 86.5%
MFM-30-12 12V 2.5A 30W 120mVp-p ±2% 90%
MFM-30-15 15V 2A 30W 120mVp-p ±2% 89%
MFM-30-24 24V 1.3A 31.2W 200mVp-p ±2% 90%
MFM-30-48 48V 0.63A 30.2W 200mVp-p ±2% 91%
 