E-Star Power Supply Mail to Us power supply products
| AC DC Power Supply | | Medical Power Supply | | DC DC Power Supply | | AC POWER CORD | - | Contact Us | - | Home |
 
AC DC Power Supply
> Medical Power Supply
> Open Frame Type

> 15W

> MFM-15

 
 
 
 
AC DC Power Supply > Medical Power Supply > Open Frame Type > 15W > MFM-15
Power Supply MFM-15
Features : * 1.93"x0.94" compact size
* Medical safety approved (2xMOPP) according to ANSI/ AAMI ES60601-1 and IEC/EN60601-1
* Suitable for BF application with appropriate system consideration
* No load power consumption<0.075W (0.09W for 3.3V)
* Extremely low leakage current
* Wide operating temp range-40~+85°C
* EMI class B for cless II configuration
* Protections: Short circuit/ Overload/ Over voltage/ Over temperature
* No minimum load required
* 3 years warranty
Model No. MFM-15
AC input voltage range 80~264VAC
Frequency Range 47~440Hz
AC inrush current (max.) Cold start 20A/115VAC 45A/230VAC
Over Load Protection 110%~150% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is removed
Leakage current Touch current <80uA/264VAC
Over voltage protection 3.8~5V, 5.75~6.8V, 13.8~16.2V, 17.3~20.3V, 27.6~32.4V
Protection type: Shut off o/p voltage, clamping by zener diode
Working temperature -35°C ~ +85°C (refer to output derating curve)
Safety standards IEC60601-1, EN60601-1, EAC TP TC 004, UL ANSI/ AAMI ES60601-1(3.1 version), CAN/CSA--C22 3rd Edition approved; Design refer to EN60335-1
EMC Emission Parameter Standard Test Level / Note
Conducted EN55011 (CISPR11) Class B
Radiated EN55011 (CISPR11) Class B
Harmonic Current EN61000-3-2 Class A
Voltage Flicker EN61000-3-3 -----
EMC Immunity EN60601-1-2
Parameter Standard Test Level / Note
ESD EN61000-4-2 Level 4, 15KV air; Level 4, 8KV contact
RF field susceptibility EN61000-4-3 Level 3, 10V/m(80MHz~2.7GHz)
Table 9, 9~28V/m(385MHz~5.78GHz)
EFT bursts EN61000-4-4 Level 3, 2KV
Surge susceptibility EN61000-4-5 Level 3, 1KV/Line-line
Conducted susceptibility EN61000-4-6 Level 3, 10V
Magnetic field immunity EN61000-4-8 Level 4, 30A/m
Voltage dip, interruption EN61000-4-11 100% dip 1 periods, 30% dip25periods, 100% interruptions 250 periods
Withstand voltage I/P-O/P: 4.0KVAC
MTBF 1210Khrs min. MIL-HDBK-217F(25°C)
Dimension 49*23.8*23mm(L*W*H)
Packing 0.028kg; 200pcs / 6.6Kg / 0.97CUFT
MFM-15 / 15W High Reliable Green Medical On Board Type
Model No. DC Voltage Rated current Rated power R & N Voltage Tol. Effi.
MFM-15-3.3 3.3V 3.5A 11.6W 150mVp-p ±1.5% 83.5%
MFM-15-5 5V 3A 15W 150mVp-p ±1.5% 85.5%
MFM-15-12 12V 1.25A 15W 150mVp-p ±1.5% 86.5%
MFM-15-15 15V 1A 15W 180mVp-p ±1.5% 87%
MFM-15-24 24V 0.63A 15.1W 180mVp-p ±1.5% 86.5%
 