Features :
* Wide Operating Voltage, 80 to 275 VAC, 47 to 63 Hz
* Single Output
* Over Load protection
* Input to Output : 2MOPP
* Size : 1”x4”x1.28”
* High ESD immunity
* Suitable professional healthcare facility
* 3 year warranty
Model No.
HBU50
AC input voltage range
Detail to see Fig.1 (Derate linearly from 100% load at 90VAC to 80% load at 80VAC)---80~275VAC
Input Frequency
Sine wave---47~63Hz
Over Load Protection
Recovers automatically after fault condition is removed---110~150%
Efficiency
Full Load, Vin=230VAC, Detail to see Rating Chart
EMC Emission
Compliance to EN55011 (CISPR11), EN60601-1-2
Operating Temperature
Detail to see Fig.2 (Derate linearly from 100% load at 40°C to 50% load at 70°C)--- -10~70°C
Storage Temperature
10 ~ 95% RH--- -40~85°C
Electro Static Discharge (ESDa)
Air Discharge, IEC61000-4-2
Electro Static Discharge (ESDc)
Contact Discharge, IEC61000-4-2
MTBF
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
Safety
IEC60601-1 Edition3.1, ES60601-1:2005(R2012), CSAC22.2 NO.60601-1:14, EN60601-1:2006/A1:2013
Model No.
Setting Voltage Range (VDC)
Output Current (A)
Maximum
Output Power (W)
Tol.
R & N (mVp-p)
Effi. (%)
min
max
min
max
min
max
HBU50-102
5.0
6.0
5
6
30
±5%
50
60
79
HBU50-103
6.0
8.0
5
6.66
40
±5%
60
80
80
HBU50-104
8.0
11.0
4.09
5
45
±5%
80
110
85
HBU50-105
11.0
13.0
3.84
4.17
50
±5%
110
130
87
HBU50-106
13.0
16.0
3.12
3.84
50
±5%
130
160
87
HBU50-107
16.0
21.0
2.38
3.12
50
±5%
160
210
88
HBU50-108
21.0
27.0
1.85
2.38
50
±3%
210
270
88
HBU50-109
27.0
33.0
1.51
1.85
50
±3%
270
330
88
HBU50-110
33.0
40.0
1.25
1.51
50
±3%
330
400
88
HBU50-111
40.0
50.0
1
1.25
50
±3%
400
500
88
Download PDF Here
< back >
--- E-Star Power Development Co., Ltd.