Line Regulation |
+/-0.5% maximum at full
load |
Efficiency |
>= 85% (82% min. for
PMP120-12, PMP120-13) |
Operating Temperature |
0°C to +60°C, Derate
from 100% at +40°C
linearly to 50% at +60°C |
Safety |
UL 60601-1, CSA C22.2 No. 601.1
TUV EN60601-1 |
EMC |
EMC performance (EN60601-1-2: 2001)
EN55011: Class B conducted, Class B radiated
FCC: Class B conducted, Class B radiated
VCCI: Class B conducted, Class B radiated
EN61000-3-2: Harmonic distortion , Class A and D
EN61000-3-3: Line flicker
EN61000-4-2: ESD,+/- 8KV air and +/- 6KV contact
EN61000-4-3: Radiated immunity, 3V/m
EN61000-4-4: Fast transient/burst, +/- 2KV
EN61000-4-5: Surge, +/- 1KV diff., +/- 2KV com.
EN61000-4-6: Conducted immunity, 3Vrms
EN61000-4-8: Magnetic field immunity, 3A/m
EN61000-4-11: Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95% reduction for
10ms |
MTBF |
150,000 hours at full load
at 25°C ambient,
calculated per MIL-HDBK-217F |