AC input voltage
range |
80 ~ 264VAC, 113 ~ 370VDC |
AC inrush current (max.) |
COLD START 30A/115VAC 65A/230VAC |
Over Load Protection |
115% ~ 160% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is removed |
Leakage current |
Touch current <80uA/264VAC |
Over voltage protection |
12.6 ~ 16.2V, 15.8 ~ 20.3V, 25.2 ~ 32.4V, 50.4 ~ 64.8V
Protection type: Shut off o/p voltage, re-power on to recover |
Working temperature |
-30 ~ +80℃(Refer to "Derating Curve") |
Safety standards |
IEC60601-1, EN60601-1, EAC TP TC 004,UL ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved; Design refer to EN60335-1(by request) |
ESD |
EN61000-4-2 |
Level 4, 15KV air; Level 4, 8KV contact |
RF field susceptibility |
EN61000-4-3 |
Level 3, 10V/m(80MHz~2.7GHz)
Table 9, 9~28V/m(385MHz~5.78GHz) |
Voltage dip, interruption |
EN61000-4-11 |
>95% dip 0.5 periods, 30% dip 25 periods,
>95% interruptions 250 periods |
MTBF |
570.5Khrs min. MIL-HDBK-217F (25℃); 1514.81Khrs min. Telcordia TR/SR-332 (Bellcore) (25℃) |
Dimension |
PCB mounting style : 87*52*29.5mm (L*W*H)
Screw terminal style : 109*52*33.5mm (L*W*H) |
Packing |
PCB mounting style : 0.197Kg;60pcs/12.8Kg/0.94CUFT
Screw terminal style :0.219Kg;50pcs/12Kg/0.56CUFT |