E-Star Power Supply Mail to Us power supply products
| AC DC Power Supply | | Medical Power Supply | | DC DC Power Supply | | AC POWER CORD | - | Contact Us | - | Home |
 
AC DC Power Supply
> Medical Power Supply
> Module Type

> 45W

> MPM-45

 
 
 
 
AC DC Power Supply > Medical Power Supply > Module Type > 45W > MPM-45
Power Supply MPM-45
Features : * 3.43" x 2.05"compact size
* PCB chassis or screw terminal mounting version
* Medical safety approved (2xMOPP) according to ANSI/ AAMI ES60601-1 and IEC/EN60601-1
* Suitable for BF application with appropriate system consideration
* No load power consumption<0.1
* Extremely low leakage current
* Wide operating temp range -30~+80°C
* EMI class B without additional components
* Isolation Class II
* Protections: Short circuit/ Overload/ Over voltage/ Over temperature
* No minimum load required
* Operating altitude up to 4000 meters (Note.7)
* 50W peak (10 sec.)
* 3 years warranty
Model No. MPM-45
AC input voltage range 80 ~ 264VAC, 113 ~ 370VDC
Frequency Range 47~63Hz
AC inrush current (max.) COLD START 30A/115VAC 60A/230VAC
Over Load Protection 115% ~ 135% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is removed
Leakage current Touch current <80uA/264VAC
Over voltage protection 5.3 ~ 7.2V, 12.6 ~ 16.2V, 15.8 ~ 20.3V, 25.2 ~ 32.4V, 50.4 ~ 64.8V
Protection type: Shut off o/p voltage, re-power on to recover
Working temperature -30 ~ +80℃(Refer to "Derating Curve")
Safety standards EAC TP TC 004, IEC60601-1, EN60601-1, UL ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved; Design refer to EN60335-1(by request)
EMC Emission Parameter Standard Test Level / Note
Conducted EN55011 (CISPR11) Class B
Radiated EN55011 (CISPR11) Class B
Harmonic Current EN61000-3-2 Class A
Voltage Flicker EN61000-3-3 -----
EMC Immunity EN60601-1-2
Parameter Standard Test Level / Note
ESD EN61000-4-2 Level 4, 15KV air; Level 4, 8KV contact
RF field susceptibility EN61000-4-3 Level 3, 10V/m(80MHz~2.7GHz)
Table 9, 9~28V/m(385MHz~5.78GHz)
EFT bursts EN61000-4-4 Level 3, 2KV
Surge susceptibility EN61000-4-5 Level 3, 1KV/Line-line
Conducted susceptibility EN61000-4-6 Level 3, 10V
Magnetic field immunity EN61000-4-8 Level 4, 30A/m
Voltage dip, interruption EN61000-4-11 >95% dip 0.5 periods, 30% dip 25 periods,
>95% interruptions 250 periods
Withstand voltage I/P-O/P: 4.0KVAC
MTBF 563.44Khrs min. MIL-HDBK-217F (25℃); 1530.14Khrs min. Telcordia TR/SR-332 (Bellcore) (25℃)
Dimension PCB mounting style : 87*52*29.5mm (L*W*H)
Screw terminal style : 109*52*33.5mm (L*W*H)
Packing PCB mounting style : 0.185Kg;60pcs/12.1Kg/0.94CUFT
Screw terminal style :0.206Kg;50pcs/11.3Kg/0.56CUFT
45W AC-DC High Reliable PCB-Mount Green Medical Power Module MPM-45 series
Model No. DC Voltage Rated current PEAK CURRENT Rated power PEAK LOAD(10sec.) R & N Voltage Tol. Effi.
MPM-45-5[] 5V 8A 8.8A 40W 44W 80mVp-p ±2% 88%
MPM-45-12[] 12V 3.75A 4.13A 45W 49.5W 120mVp-p ±2% 91.5%
MPM-45-15[] 15V 3A 3.3A 45W 49.5W 120mVp-p ±2% 92.5%
MPM-45-24[] 24V 1.88A 2.1A 45.1W 50.4W 200mVp-p ±2% 92.5%
MPM-45-48[] 48V 0.94A 1.05A 45.1W 50.4W 240mVp-p ±2% 92%
 